• 利来w66

    CO analyzer

    Model: aqms-3000 carbon monoxide analyzer 

    (1) Technical principles 

    The basic working principle of carbon monoxide analyzer is gas filtration correlation non dispersive infrared method: Based on the absorption of infrared light wave by CO molecule, it conforms to beer Lambert law,

    clip_image001.png

    The infrared light with a specific wavelength is preferentially modulated, and then passes through a gas chamber containing high concentration of carbon monoxide and a pure nitrogen gas chamber respectively. When the red light passes through a high concentration of carbon monoxide, the infrared light enters the sample gas chamber, and the co no longer absorbs the infrared light; when it passes through a pure nitrogen gas, the infrared light can be absorbed by the CO in the sample gas, and the information of the sample gas CO concentration is recorded. The measurement signal M and the reference signal R are obtained by the demodulation circuit, and the CO concentration can be calculated according to the M / R.

    clip_image002.jpg



    (2)Technical index

    指标

    数值

    测定范围

    0 ~1ppm、0~1,000ppm(可调试)

    单位

    ppb,ppm,ug/m3,mg/m3

    零点噪声

    <0.02ppm(RMS)

    最低检出限(LDL)

    0.04ppm(RMS)

    零点漂移

    <0.1ppm/24小时

    量程漂移

    <0.5ppm/24小时

    响应时间

    <120s

    线性

    <1%/FS

    重复性

    <1%

    采样流量范围

    800cm3/min±10%

    环境温度

    5℃~40℃

    尺寸(H×W×D)

    178mm×432mm×613mm

    重量

    <16kg

    电源

    200V~240V、50/60Hz、400W

    模拟输出

    10V、5V、1V、0.1V(可选)

    通讯规格

    RS232、RS485


    (3)Functional characteristics

    ■ use of gas filter related technology 

    ■ light chamber with temperature control and vibration protection 

    ■ built in AC mechanical vacuum pump 

    ■ multi function operation by microprocessor 

    ■ bidirectional RS-232 port for remote control

     ■ digital status output instrument working parameters 

    ■ with temperature compensation and pressure compensation options 

    ■ task software allows monitoring of test data during operation 

    ■ stable flow control system 

    ■ use electronic flow sensor to monitor sample gas flow 

    ■ continuous self check with alarm function 

    ■ self check function of light source intensity attenuation 

    ■ adaptive signal filtering technology to optimize response time

    ■ automatic monitoring of instrument parameters

    ■ color display, touch screen operation

    ■ large capacity memory, automatic storage of historical data

    利来w66